"PROCESS PARTICLE COUNTER
(PPC) SENSOR/CONTROLLER FOR OPTIMIZING POWER RECOVERY EXPANDER AND GAS
TURBINE PERFORMANCE"
|
|
|
APPLICATIONS NOTE FOR MEASUREMENTS
AT THE ENTRANCE AND EXIT OF A THIRD STAGE SEPARATOR |
|
|
INTRODUCTION: Process Metrix LLC, (PMC) has developed a
real-time laser-optical process particle counter/sizer (PPC) as a
short term or automated long-term sensor and control system for dust
monitoring of expanders/gas turbines and process stacks. This note
summarizes results obtained in two series of tests at a petroleum
refinery in February and July of 2002. These measurements were
obtained at the entrances of the third stage separator (TSS inlet) and
expander (TSS outlet) respectively. We used a prototype PPC instrument
to obtain these measurements, confirming the ability of the PPC
concept to accurately measure under normal system gas flow conditions
(T=1200F, 47 psia). This measurement series also confirmed the
required operating parameters for the fixed alignment PPC that is
currently being designed.
DESCRIPTION OF THE INSTRUMENT AND
EXPERIMENTAL SETUP
INSTRUMENTATION: PMC used our research PPC instrument
design, which measures over the nominal size range of 0.3 to 100
microns, at number concentrations approaching 106 particles per cm3.
The normal 1.25" by 3.5" rectangular "flow window" opening was
modified by using an external water-cooled clamshell arrangement to
accept the 0.75" dia. extraction flow line provided at the refinery.
For the majority of measurements, the PPC was inserted in the normal
line used by the refinery for mass sampling measurements, as shown in
the photograph of Figure 1. The instrument was calibrated with our
standard glass reticle (RS-2), and total mass balances were verified
using a cement powder in our laboratory wind tunnel.
SETUP: Following safety training, and completion of the
standard mass sample measurements by the refinery, PMC set up the PPC
in less than an hour after finding the proper connection hardware for
water-cooling and air purge connections.
MEASUREMENTS: After preliminary measurements to determine
appropriate instrument settings, we settled on sample times of
approximately three minutes to obtain several million particle size
measurements for the entire instrument range of 0.3 to 100 microns.
Real time display of the results is available in 3 minutes following
each measurement. Particle velocities are also obtained by measuring
the particle pulse width as the particle traverses a known laser beam
diameter. With these measurements we are able to obtain the detailed
size distribution and absolute concentration at actual flow conditions
in the probe. We monitored the temperature and pressure conditions at
the flow window, allowing determination of absolute particle loadings
in the system flow or at NTP conditions for comparison with
independent measurements of particle loading. For the July
measurements, a throttling valve was placed downstream of the PPC, so
that measurement pressures were similar to that of the primary flow
(40 psia), with temperatures of approximately 700F.
|
|

|
|
FIGURE 1. PHOTOGRAPH OF PMC PPC AND MASS SAMPLING POT
USING STANDARD SAMPLING PORTS AT THE
ENTRANCE AND EXIT OF THE TSS. |
EXPERIMENTAL RESULTS: Figure 2 shows the July PPC results
overlaid for representative upstream and downstream conditions of the
TSS. The bold solid lines show representative average frequency
distributions (and cumulatives with same colors) at the TSS inlet and
outlet at nominally iso-kinetic sampling conditions. The July results
are essentially identical to the February measurements. Figure 3 shows
Horiba ensemble diffraction measurements (provided
by the refiner) of a representative collected sample, giving the inlet
and outlet distributions of the TSS. Note that the Horiba measurements
are normalized measurements, and do not give the absolute particle
concentration. Nevertheless, the two independent measurements show
good agreement for the sizes of the distribution bi-modal peaks, at 1
and 28 microns respectively. The TSS essentially removes the larger
end of the distribution, with little impact on the nominally one
micron peak. This is consistent with the design cut-point of the TSS,
which is approximately three microns. Note that the PPC shows more
detail than the Horiba results (which show no particles above 3
microns) for the 10% of the mass that is above the TSS cut-point,
providing more detailed information about the critical larger
particles entering the expander.
|
|

|
|
FIGURE 2. PPC FREQUENCY AND CUMULATIVE
SIZE DISTRIBUTION RESULTS (NOT NORMALIZED) FOR
COMPARISON WITH HORIBA INSTRUMENT RESULTS OF FIGURE 3. CONCENTRATION UNITS
ARE VOLUME OF
SOLIDS PER VOLUME OF GAS AT MEASURED TEMPERATURE AND PRESSURE. |
|

|
|
FIGURE 3: MEASUREMENTS (BY HORIBA,
ENSEMBLE DIFFRACTION) OF COLLECTED SAMPLES AT THE TSS INLET AND OUTLET. |
MASS CONCENTRATIONS: The next level of comparison is to
determine the absolute total mass measured by the sample collection
system compared with the equivalent integrated mass obtained by the
PPC. This requires that we know the particle envelope density for
direct mass comparison along with the gas density, determined by the
temperature and pressure measurements at the probe. Incorporating this
information, Table 1 summarizes the comparisons of the average total
mass and ratios of the mass, upstream/downstream of the TSS. The
ratios of the two techniques are within 5%, although the PPC total
mass results average about 25% higher than the method 5 filter
measurements. The variances of the filter measurements range from
5-15%, and the variances of the PPC results are on the order of 5%. We
believe the envelope density characterization is within 5%, based on a
liquid displacement measurement, which does not penetrate the pore
structure of the catalyst substrate. These uncertainties do not close
the entire measurement value gap (approximately 15-20% additional
difference), so there are probably systematic errors in the size,
particle shape, cross-section, and velocity measurements of the PPC.
There could also be some systematic transport losses in the method 5
filter measurements. Nevertheless, these results show promising
comparisons between an optical method and the traditional method 5.
|
|

|
Figure 4 shows the important capability of the PPC to measure
essentially continuously, providing time variation information
downstream of the TSS. A complete size distribution and mass loading
was obtained at approximately 6 minute intervals over a period of 4
hours. For the comparable period, two filter measurements were taken
at the beginning and end of this 4 hour period, showing similar
increases (25-30%) in total mass for the days of July 18, 19. For July
17, there was essentially no change in the mass loading measured by
the PPC or the filter collector.
|
|

|
|
FIGURE 4. VARIATION OF CATALYST
LOADING AND PARTICLE SIZE AS A FUNCTION OF RUN NUMBER OVER A
PERIOD OF APPROXIMATELY 4 HOURS. |
MEASUREMENT SUMMARY
- Size distribution measurements with the PPC are in good
agreement with independent measurements of mass samples both
upstream and downstream of the TSS.
- Absolute total mass measurement comparisons are within 25% of
each other for both TSS inlet and outlet measurements and ratios of
inlet/outlet are within 5%. Outside of random variations, remaining
differences in mass closure are most likely due to systematic errors
of the PPC and Method 5.
FIXED ALIGNMENT PPC: Although the basic optical configuration
of the PPC will be similar to the current research system, the
instrument is being completely updated with new, faster, and more
accurate (temperature corrected) electronics, Windows real time
software, and a fixed alignment optical system. The flow window
design will change to include the option for an in situ measurement
(for low pressure systems) or for a sampling system approach for
high pressure applications. The new design will be process
compatible without operator intervention (after initial installation
and startup). The new design will be available for sale by
November, 2002
|
FURTHER INFORMATION: Interim testing capabilities with the
PPC research instrument supported by PMC engineers are currently
available. For more details on the feasibility of PPC measurements in
your process, please contact:
|
|
DON HOLVE
PROCESS METRIX LLC
2110 OMEGA RD. STE. D
SAN RAMON, CA 94583
T: 925 837 1330
F: 925 837 3864
E: DHOLVE@PROCESSMETRIX.COM
WEB: WWW.PROCESSMETRIX.COM |
The principals of PMC have successfully developed and sold two
other particle sizing instrument technologies, including the Insitec,
now manufactured by Malvern Instruments, for continuous process
control in powder milling applications. PMC is currently developing
the HiVol instrument for Inficon Corp. for applications in the
semi-conductor industry. The PMC website provides further details on
other process instrumentation developments for the Steel Industry.
PMC has over 45 man-years of experience in developing and applying
real-time optical instrumentation for harsh industrial environments.
ACKNOWLEDGEMENTS: This development work is being supported by three
contracts. The first is from the DOE Office of Industrial
Technologies, supporting Pulp and Paper applications, and the second
one is directed specifically for expanders/gas turbine applications
funded by the Inventions and Innovation Office. A third DOE SBIR has
just been received and will focus on measurements in high pressure
gasifiers. FUTURE APPLICATIONS DEVELOPMENT WORK: Process Metrix has
submitted a 50% cost share proposal to DOE in June, 2002 for further
applications development of the PPC. Another proposal to DOE SBIR will
be submitted August 14, 2002 for development of a method 5 equivalent
optical instrument. In-kind cost sharing credit for use of industrial
facilities is solicited by PMC in conjunction with these proposals.
Effectively, this will allow the user to test the PPC at no direct
cost or risk. Industrial support is the key to a successful proposal
effort. DOE contract cost-sharing participants include:
WEYERHAEUSER CORPORATION
INTERNATIONAL PAPER CO.
DRESSER-RAND CORPORATION
MEGGITT AVIONICS
SHELL GLOBAL SOLUTIONS (US)
SIEMENS/WESTINGHOUSE
|
Close
|